JCSMR/TSRI Sort Recovery Page

This page thanks to Joe Trotter "Sort Guru"

Version 2.0 
Modified for both normal BDIS 0.25 drop clock cycle and the modified TurboSort ® 0.1 drop clock cycle
Input Parameters Output Parameters 
Sorter: Standard TurboSort ®  Deadtime Loss 
Mode: Normal-R Count  Throughput hr 
Sort Envelope: drops + clock cycles Sort Left 
Sample Flow Rate: events/sec  Rate /hr /sec 
Frequency: Hz  Yield hrmin
Max loss: % of sort left subset Max: / sec
Pulse Width: uSec  drops/event 
Dead Time: uSec  Sort Right 
Sort fractions Left:% Right:  Rate / hr /sec 
Yield Left: Right:  Yieldhrmin
Action:
  

 

These calculations apply to standard BDIS sorters where the instrument is set to trigger on light scatter. These calculations assume ideal conditions and that there are no other hidden dead times wihtin the sorting electronics. In practice, there are internal constraints within some sort circuits that will cause additional aborts that may or may not be counted by the electronics. For High Speed Enrichment, everything changes based on signal/noise and other sample characteristics. Your milage may vary... always check your instruments efficiency...


03/06/98

Comments & Questions: trotter@scripps.edu