Instruments

Equipment

LSR II

For more information about the LSR II click here.
For more information about the HTS attachment click here
To download a step by step manual to run plates on the HTS click here or download the Daily Operation Tutorial

 


 

The BD - LSR Benchtop Flow Cytometer

Click on the image for more information about the LSR.


Vantage DiVa:

More information coming soon!




FACStar plus:

FACStar plusThe FACStar plus is equipped with a triple laser bench, Macrosort Sort Option (for large particles), Sort Enhancement Module (allows nonrectangular sort windows), Pulse Processor (Ratio Measurements for Calcium Studies, Doublet Discrimination for DNA Analysis) and an Automatic Cell Deposition Unit.

The ACDU is used to sort one or more cells per well into plates from Terasaki to 24 well plates or slides.

There is a cooling/heating unit attached which allows temperature control from 4°C - 56°C.

 

 

 

 

 

 

 

 

 

 

Laser 1 is a Coherent Argon-Ion water cooled laser tuned to 488nm wavelength.

Laser 2 is a Spectra Physics Argon-Ion water cooled laser tuned to UV.

Laser 3 is a Helium-Neon air cooled laser at 632nm wavelength.

 

 

 


 

BD FACSCalibur:

The FACSCalibur is an analytical flow cytometer with two scattered light detectors and four fluorescence detectors. The fourth fluorescence detector option which this instrument has collects signals from fluorochromes excited by a small 633nm red diode laser.

Choose this cytometer in preference to the BD LSR, if you are working with APC and a tandem dye such as PE-CY5 as in practice you will have less difficulty with the compensation for spectral overlap on this instrument.

The FACSCalibur is also fitted with the Multiwell Acquistion System with allows samples to be taken directly from 96 and 384 well plates

If you'd like to use this unit, please read the MAS User Guide PDF file and remember to remove the MAS probe .


FACSort:

The FACSort can measure up to six parameters. FSC, SSC and four fluorescence parameters. It can also measure the pulse height, area and width of three fluoresence parameters (FL1, FL2 and FL3).

 

 

 

 

 

 

 

 

Some of the most common fluoresent dye's used on FACSort here at the JCSMR.

Fluorescent Dye Filter Peak Emmision Detector
FITC ( Fluorescin Isothiocyanate) 530nm/30 525nm FL1
Phycoerythrin 585nm/42 578nm FL2
Propidium Iodide 585nm/42 or 670LP 617nm FL2 / FL3

PerCp (Peridinin chlorphyll protein)

670LP 675nm FL3
Cychrome/Tri-colour (PE-Cy5) 670LP 670nm FL3
APC 661nm/16 660nm FL4
Alexa 647 661nm/16 668nm FL4

 

The flow rates on the FACSort are: LO ( 12µl ±3µl /min ) MED (35µl ±5µl /min) HI (60µl ±7µl/min)

 


FACScan:

Facscan
Click here to go to start up and shut down protocols for the FACScans. A different version is also available as a PDF file.

The FACScan is a single air cooled laser (488nm) instruments and are capable of three colour and two scatter parameter analysis. The FACScan is also equipped with a Doublet Discrimination Module for the measurement of pulse width and area which is used in DNA analysis.

Some of the most common dye's used on FACScan's here at the JCSMR.

Fluorescent Dye Peak Emmision Detector
FITC ( Fluorescin Isothiocyanate) 525nm FL1
Phycoerythrin 578nm FL2
Propidium Iodide 617nm FL2 / FL3

PerCp (Peridinin chlorphyll protein)

675nm FL3
Cychrome/Tri-colour (PE-Cy5) 670nm FL3

For more information about fluorescence spectra's go to Joe Trotters Spectrum Viewer (Java) at the Scripps Research Institute.

 

Updated last 1-8-2005 by Sabine Grüninger

 